E45-282 · Calibration · Humidity Sensor · ★
Calibrated in a 30℃ humid environment; on site at 20℃ dry the reading was off by 5%RH, with poor consistency
[Mechanism] Humidity-sensing elements are cross-sensitive to temperature and humidity; a large difference between the calibration environment and the measured environment makes the calibration curve unrepresentative of the site, and the measurement shifts with the operating conditions.
[Calculation] A 10℃ temperature difference introduces a 2%RH error and the humidity difference another 3%RH, a total of 5%RH, exceeding the 2% tolerance.
[Design Omission] Calibration was not done under representative conditions and no temperature/humidity compensation was applied.
① Near operating conditions: calibrate close to the site temperature/humidity → representative curve
② Temperature compensation: build a temperature compensation table → remove the temperature error
③ Humidity compensation: add a humidity compensation term → remove the humidity error
④ Multi-point: calibrate at different temperature/humidity points → accurate over the whole range
⑤ Recheck: compare on site → confirm agreement
⑥ Out-of-tolerance alarm: alarm when the error >2% → early warning
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the environment requirement into the procedure → control at the source
JJG 875-2019 Verification Regulation of Digital Thermo-Hygrometers