Knowledge Base › Electrical Sensor Error Case › Digital Filter Left On Distorting Calibration

Digital Filter Left On Distorting Calibration

E45-283 · Calibration · Vibration Accelerometer · ★

Problem Description

The digital filter was not disabled during calibration; smoothing made the peak read low by 12% and the sensitivity was set wrong

Root Cause

[Mechanism] Leaving the digital filter on during calibration smooths the transient excitation signal, so the measured peak falls below the true value and the sensitivity coefficient derived from the low peak is too small.
[Calculation] A 10ms filter time constant smooths a 2ms shock peak by 12%; the sensitivity is calibrated to 0.88 times, making readings low across the range.
[Design Omission] Calibration did not require the filter to be off, and a steady-state sine excitation was not used.

Corrective Solution

① Disable the filter: turn off the digital filter during calibration → undistorted peak
② Sine calibration: calibrate with a steady-state sine → accurate steady amplitude
③ Re-enable after: turn the filter back on after calibration → stable operation
④ Recheck: verify on a shock table after calibration → verify sensitivity
⑤ Procedure: write the filter-off requirement into the calibration procedure → prevent omissions
⑥ Out-of-tolerance alarm: alarm when the sensitivity error >5% → early warning
⑦ Records: keep the calibration state on file → traceability
⑧ Write the process into the specification → control at the source