E45-290 · Calibration · Load Cell · ★
No repeatability test was run; three readings with the same weight differed by 0.3%FS, causing batch-filling fluctuation
[Mechanism] Calibration was done only once, without verifying repeatability, so random error and loose mounting went unnoticed and readings scatter under the same load.
[Calculation] Three readings at the same load differed by 0.1/0.2/0.3%FS; the repeatability of 0.3%FS exceeds the 0.1% limit.
[Design Omission] No repeatability test was performed and the scatter was not evaluated.
① Repeatability test: measure three times at the same load → verify the scatter
② Scatter check: repeatability <0.1% → stable
③ Re-mounting: if out of tolerance, re-mount → secure
④ Recheck: re-verify at different loads → full range
⑤ Out-of-tolerance alarm: alarm on large scatter → early warning
⑥ Records: keep the repeatability data on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the repeatability requirement into the procedure → control at the source
JJG 669-2003 Verification Regulation of Load Cells