E45-293 · Calibration and Adjustment · Load Cell · ★
A sensor of the same model is replaced without recalibration; the new unit's zero is off by 1%FS and the batching differs
[Mechanism] Same-model sensors differ in individual zero offset and sensitivity coefficient; replacing one while keeping the old calibration coefficients makes the new unit deviate from its true characteristic.
[Calculation] The new unit's zero offset is 1%FS; keeping the old coefficients biases the whole range by 1% and the batching goes out of tolerance.
[Design Omission] Recalibration on replacement was not mandated and individual coefficients were not stored.
① Recalibrate on replacement: recalibrate whenever a unit is replaced → coefficients matched to the unit
② Store individual data: store each unit's calibration coefficients → for matching
③ Verification: compare against an actual weighing after calibration → verify
④ Out-of-tolerance alarm: warn when the deviation >0.5% → early warning
⑤ Records: keep the replacement on file → traceability
⑥ Periodic calibration: add a calibration after replacement → prevent drift
⑦ Pairing selection: select and pair multiple units → consistency
⑧ Write the replacement rules into the system → control at the source
JJG 669-2003 Verification Regulation of Load Cells