Knowledge Base › Electrical Sensor Error Case › Wrong Auto-Calibration Script

Wrong Auto-Calibration Script

E45-294 · Calibration and Adjustment · Temperature Transmitter · ★★

Problem Description

The auto-calibration script has a wrong formula; after batch calibration everything is off by 3%FS

Root Cause

[Mechanism] The auto-calibration program has a wrong fitting formula or unit conversion, so batch calibration writes a uniformly wrong coefficient and the whole batch of sensors carries a systematic bias.
[Calculation] The script mistakes ℃ for K, an error of 273; the wrong conversion biases the whole range by 3%FS and the batch is distorted.
[Design Omission] The script was not reviewed and no manual verification on a sample was done.

Corrective Solution

① Script review: two-person review of the calibration script → correct formula
② Sample check: manually calibrate a sample and compare first → verify the script
③ Unit check: check ℃/K/unit consistency → no mixing
④ Verification: sample from the batch → confirm consistency
⑤ Error alarm: warn when the deviation >1% → early warning
⑥ Rollback: roll back and recalibrate on any anomaly → correct the bias
⑦ Records: keep the script version on file → traceability
⑧ Write the script rules into the system → control at the source

Checklist

Reference

JJF 1001-2011 General Terms in Metrology and Their Definitions