Knowledge Base › Electrical Sensor Error Case › Inaccurate Calibration Weights

Inaccurate Calibration Weights

E45-295 · Calibration and Adjustment · Load Cell · ★

Problem Description

Calibrating with corroded weights leaves a 1.2%FS deviation, and batching is systematically short

Root Cause

[Mechanism] The standard weights were not verified for a long time and rusted, losing mass; as the calibration reference their mass is low, so the sensor under test is calibrated with a wrong mass coefficient.
[Calculation] A weight nominal 10kg actually weighs 9.9kg, 1% short; after calibration the coefficient is 1% low and the whole range reads 1.2% low.
[Design Omission] The weights were not checked and were not sent for periodic verification.

Corrective Solution

① Weight check: check the weight certificate before calibration → true reference
② Periodic submission: annual verification of the weights → no rust loss
③ Grade matching: the weight grade is three times that of the unit under test → true value
④ Verification: recalibrate with another set of weights → confirm consistency
⑤ Out-of-tolerance alarm: forbid calibration with out-of-tolerance weights → prevent errors
⑥ Certificates: keep the certificate numbers → traceability
⑦ Storage: store moisture-proof and rust-proof → long-term
⑧ Write the weight management into the system → control at the source