Applications: precise measurement of outside diameter, thickness and length
Accuracy: 0.01mm (ordinary type) / 0.001mm (digital type)
Range: 0~25mm, 25~50mm, 50~75mm... (in 25mm steps)
Screw micrometer principle: the micrometer screw has a pitch of 0.5mm, so one revolution of the thimble moves 0.5mm
The thimble is divided into 50 divisions, each division = 0.5/50=0.01mm
① Read the integer and 0.5mm on the sleeve: look at the sleeve scale just to the left of the thimble edge
② Read the decimal on the thimble: find the thimble graduation aligned with the sleeve datum line
③ Measured value = sleeve reading + thimble reading
① Zero the instrument before use: check the zero position with a gauge bar
② Use the ratchet (force device): stop when a click is heard, ensuring consistent measuring force
③ The workpiece and instrument must be at the same temperature (let steel parts stand at room temperature for 30 min)
④ The workpiece axis must be perpendicular to the anvil during measurement
⑤ Never measure a workpiece that is rotating
Wipe clean and apply anti-rust oil after use, store in the dedicated case
Send for periodic metrological verification (the verification cycle is generally 1 year)
Applications: measurement of inside and outside diameters, length, depth and steps
Accuracy: 0.02mm (50 divisions), 0.05mm (10 divisions), 0.01mm (digital)
Range: 0~150mm, 0~200mm, 0~300mm are common
Main scale + vernier scale: vernier division value = main scale graduation spacing / n
For a 20-division caliper: 20 vernier divisions correspond to 19mm on the main scale, each division = 1/20=0.05mm
The actual 0.02mm accuracy is a 50-division vernier: 50 vernier divisions correspond to 49mm on the main scale, each division = 0.02mm
① Read the integer on the main scale: the main scale graduation just to the left of the vernier zero
② Read the decimal on the vernier: the aligned graduation × division value
③ Measured value = main scale integer + number of aligned vernier divisions × division value
① Close the jaws and check the zero position
② Use the lower jaws for outside diameters and the upper jaws for inside diameters
③ Use the depth rod for depth measurement
④ After the jaws contact the workpiece, tighten the lock screw before reading
⑤ Do not apply excessive force (the jaws will deform)
⑥ Do not measure blanks or moving workpieces
Wipe clean and oil, avoid knocks and drops, verify periodically
Applications: measurement of bore diameters, groove widths and form errors (roundness / cylindricity)
Accuracy: 0.01mm
Range: interchangeable heads cover ranges such as 10~18 / 18~50 / 50~160mm
Consists of a dial indicator + a lever transmission mechanism + interchangeable measuring heads
Head displacement is transmitted through the lever to the dial indicator and magnified for display
① Select an interchangeable head matching the bore diameter range
② Zero with an outside micrometer or a standard ring gauge: compress the head 0.5~1mm and set to zero
③ When measuring, insert the stem into the bore and rock it left and right to find the minimum reading (the diameter direction)
④ Rock it up and down to find the maximum reading (ensuring the head is in the diameter position)
⑤ Reading = zeroing dimension + dial indicator deviation
① Preload of 0.5~1mm is mandatory (ensures the head contacts the bore wall)
② Read at the turning point: the left-right minimum corresponds to the diameter
③ Measure several sections (top / middle / bottom) to judge cylindricity
④ Keep the head and workpiece clean so that oil films do not affect the result
The dial indicator is a precision component; avoid impact and return it to zero after use
Applications: measurement of height and depth, and scribing (marking lines on the workpiece for positioning)
Accuracy: 0.02mm (vernier type), 0.01mm (digital type)
Range: 0~300mm, 0~500mm, 0~1000mm are common
Base + column + vernier scale + measuring jaws (scribing jaw / measuring jaw)
Digital types have an LCD display and can preset values
① Place the height gauge on a surface plate, with the base bottom face in contact with the plate
② Scribing: adjust the jaw tip to the target height, lock it, then drag along the plate to mark the workpiece
③ Measuring height: bring the jaw measuring face into contact with the measured surface; the reading is the height
④ Measuring depth: fit the depth jaw, insert it into the bore or groove and read the depth
① Must be used on a granite surface plate or a cast iron surface plate
② Keep the base clean and in close contact with the plate
③ Apply even force when scribing; excessive force will damage the workpiece
④ The column must be perpendicular to the plate (check the squareness periodically)
⑤ Carry the instrument by the base, never by the column
Apply anti-rust oil to the guide rails; digital models must be re-zeroed after a battery change
Applications: measurement of the gap between two surfaces (fit clearance, guideway gap, end face gap)
Specification: thickness 0.02~1.00mm, each blade differs by 0.01mm (thin end) or 0.05mm (thick end)
Set: usually 17~21 blades form a set
A combination of thin steel blades, each marked with its thickness
Material: spring steel (good elasticity, not easily bent)
① Select a blade of appropriate thickness
② Insert the blade into the gap being measured
③ Trial insertion: the blade that can be inserted but with slight resistance gives the gap value
④ Combining blades: when one blade is insufficient they may be stacked, but do not stack more than 3
① Insert gently, never force (this will scratch the surface or break the blade)
② Blades should be clean, free of oil and free of creases
③ Remove the blades after measuring; do not leave them in the gap for long (the spring force causes deformation)
④ A bent blade must not be reused (its thickness is no longer accurate)
⑤ When blades are stacked, the gap value = the sum of the blade thicknesses
Piston to cylinder clearance, bearing clearance, guideway gib clearance, mould parting face clearance
Wipe clean and apply anti-rust oil after use, fold into the dedicated case
Applications: comprehensive inspection of whether the pitch diameter of a thread is within tolerance (specific values cannot be measured, only pass or fail)
Types: cylindrical thread gauges (ring gauge / plug gauge for external threads, plug gauge for internal threads)
Simulates the maximum material boundary and controls the effective pitch diameter not exceeding the maximum value
The GO gauge should screw in smoothly over the full length of the thread under test
If it cannot be screwed in → effective pitch diameter too large → not acceptable
Simulates the least material boundary and controls the actual pitch diameter not falling below the minimum value
The NO-GO gauge must not screw in more than 2 pitches
If it screws in more than 2 threads → actual pitch diameter too small → not acceptable
Acceptance condition: GO gauge passes + NO-GO gauge does not pass
① Clean the thread and the gauge, removing burrs and chips
② GO gauge: screw in by hand; it should pass freely over the full length (without force)
③ NO-GO gauge: screw in from both ends; no more than 2 threads is acceptable
④ Judgement: GO passes + NO-GO fails = acceptable
① Never screw in forcibly (this damages the gauge or causes a wrong judgement)
② The gauge and the workpiece must be at the same temperature (large differences affect the judgement)
③ Gauges are verified periodically (wear changes the pitch diameter)
④ Thread gauges cannot measure pitch or profile errors (use the three-wire method or a thread micrometer)
Applications: measurement of surface roughness parameters (Ra/Rz/Rt/Rmax etc.)
Types: stylus type (contact), optical type (non-contact)
Accuracy: can measure the Ra range 0.005~50μm
Ra: arithmetical mean deviation of the profile (most commonly used)
Rz: maximum height of the profile (mean of the peak-to-valley sums over 5 sampling lengths)
Rt: maximum peak-to-valley height of the profile
Rmax: maximum peak-to-valley height within a single sampling length
A diamond stylus (radius 2/5/10μm) slides along the surface
The sensor picks up the vertical displacement signal → filtering → calculation of roughness parameters
① Select the parameter: Ra (general measurement) / Rz (when waviness is large)
② Select the sampling length: 0.8mm for Ra>0.1μm, 0.25mm for Ra<0.1μm
③ Evaluation length = 5× sampling length (as specified by the standard)
④ Measure with the stylus perpendicular to the machining lay direction
⑤ Place the instrument level and set the stylus lightly on the workpiece surface
① The stylus is a consumable; avoid dragging it quickly over hard surfaces
② Measure perpendicular to the machining marks (most sensitive direction)
③ The workpiece surface must be clean and free of oil
④ Calibrate periodically with a standard reference specimen
⑤ Stylus instruments cannot measure very hard or very soft surfaces (they scratch or stick)
Quick shop-floor inspection: portable stylus roughness tester
Laboratory precision analysis: optical profilometer / white light interferometer
Applications: precise measurement of three-dimensional geometry, form errors and location errors
Accuracy: 0.5~3μm (depending on model and environment)
Range: from 500×400×400 up to several meters
Moving bridge: most common, the workpiece stays put and the bridge moves
Fixed bridge: higher accuracy, suitable for small and medium parts
Cantilever: suitable for large sheet metal parts
Displacements of three orthogonal guideways are read by scales to give coordinates (X, Y, Z)
The probe contacts the measured point → the coordinate is recorded → software calculates the geometric quantities
Contact probe (ruby ball): measures dimensions, form and location
Scanning probe: continuous point acquisition for profiles
Optical / laser probe: non-contact measurement of soft or easily deformed parts
① Place the workpiece on the table and calibrate the datums (establish the part coordinate system)
② Determine the measurement strategy: number of points, measurement path
③ Probe calibration: use a standard sphere to calibrate the probe ball diameter and position
④ Automatic or manual point acquisition
⑤ The software outputs a report: dimensional deviations, form and position tolerances, GD&T evaluation
Precision measurement of complex housings / blades / gears
Reverse engineering (point cloud acquisition to rebuild the CAD model)
Constant temperature 20±2℃, humidity 55±10%, vibration <2μm, dust free
Air bearing guideways require clean compressed air (with air filtration)
Verify the accuracy periodically; handle the probe gently
Applications: precise measurement of two-dimensional geometry (length, angle, arc, hole spacing, profile)
Accuracy: 2~5μm (typical)
Features: non-contact measurement, suitable for thin / soft / small parts
Optical microscope + CCD camera + computer image processing
The X/Y stage reads coordinates from scales
Z axis with automatic focus (2.5D models have Z measurement capability)
Automatic fitting of points / lines / circles / arcs
Distance, angle and coordinate measurement
Automatic edge detection (edge extraction algorithm)
Profile scanning
① Place the workpiece on the stage and select an objective of suitable magnification
② Focus until the image is sharp
③ Automatic or manual point acquisition: the software identifies the edge automatically
④ The software calculates the geometric quantities and outputs a report
Advantages: non-contact (no scratching of the workpiece), high efficiency (automatic edge detection), suitable for small parts
Disadvantages: cannot measure in the depth direction (mainly 2D), affected by workpiece reflectivity and colour
PCB boards / mobile phone parts / connectors / springs / small stampings
Precision die inserts / micro gears
① The workpiece surface must be clean and free of reflective oil
② Choose a suitable magnification: low magnification for large parts, high magnification for fine details
③ Measure within the same focal plane, otherwise parallax affects accuracy
④ Calibrate periodically with a standard glass line scale
Applications: three-dimensional precision measurement of large workpieces and on-site assembly (meters to tens of meters)
Accuracy: ±(15μm+5μm/m) (typical, degrading with distance)
Range: spherical radius up to 20~80m
Laser interferometry (IFM) + absolute distance measurement (ADM) dual system
Two-axis rotary encoders measure the horizontal and vertical angles
Spherical measurement reflectors (SMR) reflect the laser back to the receiver
The tracker emits a laser → the reflector reflects it back → the receiver detects it
Interferometry gives the distance r, the encoders give the angles θ and φ
Measured point coordinates = (r·sinθ·cosφ, r·cosθ, r·sinθ·sinφ)
① Set the instrument on a stable tripod or tower
② Place the reflector on the measured point, held by the operator or fixed with a magnetic base
③ The instrument automatically tracks the reflector and records coordinates in real time
④ Multi-station measurement with common-point stitching (to resolve occlusion)
Large-scale assembly measurement of aircraft and spacecraft
Geometric accuracy inspection of large machine tools and construction machinery (straightness / flatness / perpendicularity)
Wind turbine blade and antenna surface profile inspection
Ship section alignment
Advantages: huge range (up to tens of meters), portable on-site measurement
Disadvantages: accuracy lower than a CMM (errors accumulate with distance), the reflector must be moved manually, susceptible to environmental disturbance (air currents / temperature)
① The reflector is extremely precise (sphericity <0.5μm); do not knock it
② Keep the measurement path free of obstruction (the laser must be re-acquired if interrupted)
③ Control the environmental temperature gradient (it affects laser ranging accuracy)
④ Verify periodically and compensate the angular encoder errors
Applications: precise measurement of form errors of revolved parts such as roundness, cylindricity, coaxiality and runout
Accuracy: roundness measurement accuracy 0.02~0.1μm
Turntable type: the workpiece rotates with the table and the probe is fixed (suitable for small parts)
Spindle type: the probe rotates with the spindle and the workpiece is fixed (suitable for large parts)
A high-precision air bearing spindle provides the rotational datum (radial runout <0.1μm)
The probe contacts the workpiece surface, and the spindle rotates through one revolution while sampling
The recorded radial variation is the roundness error profile
① Least squares circle method (LSQ): most commonly used, roundness = maximum peak-to-valley of the profile relative to the least squares circle
② Minimum circumscribed circle method (MCC): for shaft/hole fits
③ Maximum inscribed circle method (MIC): for holes
④ Minimum zone method (MZC): gives the smallest result, used for arbitration
1~15 waves per revolution: evaluate roundness (removes low-frequency eccentricity and high-frequency roughness)
1~50 / 1~150 waves per revolution: analyse the sources of high-frequency error
① Place the workpiece on the table or fixture and align it (coaxial with the spindle <0.01mm)
② Preload the probe 0.1~0.2mm, perpendicular to the measured surface
③ Select the measurement section (several sections can be measured for cylindricity)
④ After acquisition the software evaluates and outputs the roundness value and polar diagram
① Use a moderate probe force to avoid scratching
② Let the workpiece reach the same temperature before measuring
③ Multi-section measurement can determine cylindricity, taper and barreling
④ Spindle accuracy is the core element; verify periodically with a standard hemisphere
Applications: rapid assessment of the Ra value of a workpiece surface by visual and tactile comparison
Features: portable, fast, no instrument needed, suitable for preliminary shop-floor judgement
A set of standard specimens with calibrated Ra values (one group for each process such as turning / milling / grinding / boring / planing)
Each specimen is marked with: machining method + Ra value + sampling length
① Select the specimen group matching the process of the workpiece under test
② Place the specimen next to the workpiece under identical lighting
③ Visual comparison: observe surface lustre, texture and scratch density
④ Tactile comparison: run a fingernail or finger across both and compare
⑤ Ra judgement: if the workpiece appearance lies between two specimens, take the intermediate value
Rapid shop-floor inspection (first-off / patrol inspection)
Where instruments cannot be used (high temperature / vibration / oily environments)
To decide whether the part needs precision testing in a metrology room
① The specimen and the workpiece must use the same process (compare a turned part with a turned specimen)
② Specimens must be verified and calibrated periodically (wear changes the appearance)
③ For coarse judgement only; the final acceptance decision rests with the roughness tester
④ Different operators may judge differently (strong subjectivity)
⑤ Avoid rust or scratches on the specimen surface
Apply anti-rust oil after use and store in the dedicated case in Ra order
Applications: measurement of material hardness, which indirectly reflects strength and wear resistance
Types: Brinell (HB), Rockwell (HRC/HRB), Vickers (HV), Leeb (HL)
Principle: a steel ball or cemented carbide ball is pressed into the workpiece and the indentation diameter is measured to compute the hardness
Load: 3000kgf (steel ball 10mm, standard condition)
Applicable to: annealed / normalized / quenched-and-tempered parts, cast iron and non-ferrous metals (HB<450)
Advantages: a large indentation reflecting the average hardness; disadvantage: damages the surface
Principle: a 120° diamond cone is pressed in and the depth difference is measured to give a direct reading
Minor load 10kgf, total load 150kgf (HRC)
Applicable to: quenched steels / die steels / cemented carbides (HRC 20~70)
Advantages: direct reading, little damage; disadvantage: the small indentation is sensitive to local conditions
HRB (1/16 inch steel ball, 100kgf): measures soft steel and copper alloys (HRB 0~100)
Principle: a 136° diamond pyramid is pressed in and the indentation diagonals are measured to compute the hardness
Load: 1~100kgf (small loads in the laboratory measure thin layers)
Applicable to: nitrided / carburized surface layers, sheet metal, micro-hardness
Advantages: can measure thin layers, high accuracy; disadvantage: reading requires a microscope
① Select the hardness scale: HB / HRC / HV according to the material hardness and thickness
② Prepare the workpiece surface: grind flat and remove the oxide scale, Ra<0.8
③ Fit the indenter and select the load
④ Load → hold (10~15s) → unload
⑤ Read the value: directly from the dial for Rockwell, or measure the indentation and consult the table for Brinell and Vickers
① Workpiece thickness ≥8 times the indentation depth (to avoid deformation on the reverse side)
② Indentation spacing ≥4 times the indentation diameter (to avoid work hardening effects)
③ Hardness on a cylindrical surface requires a correction (curvature effect)
④ Calibrate the hardness tester periodically with standard hardness blocks
⑤ HRC to HB conversion (reference values): HRC 20≈HB 226, HRC 40≈HB 370, HRC 60≈HB 650
Related quick references: Tolerances & FitsSurface RoughnessMaterial Heat TreatmentCast Part DesignGeometrical TolerancesDimension ChainsWelded StructuresSheet Metal StructuresShafting DesignSeal DesignMaterial SelectionHeat TreatmentDatum ReferencingTolerance Principles