Ammeters go in series, voltmeters in parallel; wrong placement means short circuit or open circuit.
Range selection: estimate from nameplate, start high, keep pointer above 2/3 scale, verify polarity by touching before fixing.
Match meter impedance to the circuit: 20kΩ/V for control circuits, 10MΩ DMM for electronics; CT secondaries must never open.
| Meter | Connection | Start range | Typical error |
|---|---|---|---|
| DC ammeter | Series | Highest | Across supply - blown |
| AC ammeter+CT | Primary through window | Per ratio | Open secondary |
| DC voltmeter | Parallel | Highest | Reversed leads |
| Analog multimeter Ω | De-energized | ×1 to ×10k | Live resistance test |
| DMM voltage | Parallel | Auto/high | Current jack to volts |
| DMM current | Series | 10A→mA | Current jack across supply |
Check jack positions before and after every measurement; return the dial to the highest voltage range when done.
The quantities and component characteristics discussed here follow current national standards and the corresponding IEC documents: Instrument classes per GB/T 7676; DMMs per GB/T 13978. Always check the device datasheet and the product design file for ratings and derating.
Locate the issue with the approach above, then confirm with measurements; safety-related circuits must follow the standards and manufacturer requirements.