X caps bridge L-N (differential, fail short → fuse); Y caps go line-to-earth (fail short → shock hazard).
Y-cap size is capped by touch current: I=2πfCU; medical gear allows only ~0.1mA, hence 1~4.7nF typical.
Always certified parts (X1/X2, Y1/Y2), never generic ceramics; keep the X-cap discharge resistor.
| Aspect | X capacitor | Y capacitor |
|---|---|---|
| Connection | Across L-N | Line/N to earth |
| Filters | Differential | Common mode |
| Failure | Short → fuse | Short → live earth |
| Class | X1/X2/X3 | Y1/Y2 |
| Typical size | 0.01~1μF | 1~4.7nF |
| Constraint | Discharge resistor | Touch current |
Apply the points above to X and Y safety capacitors at the supply input; the figures are typical engineering values, so confirm them against the device datasheet and the actual design before release.
The quantities and component characteristics discussed here follow current national standards and the corresponding IEC documents: Per GB/T 14472, IEC 60384-14, GB 4943.1, GB 4706.1 and the corresponding IEC documents. Ratings and derating follow the device datasheet and the product design file. Always check the device datasheet and the product design file for ratings and derating.
Locate the issue with the approach above, then confirm with measurements; safety-related circuits must follow the standards and manufacturer requirements.