A thermocouple outputs millivolts that also shift with cold-junction temperature, while an RTD is a small resistance that varies with temperature. Both need an op amp front end. This page covers cold-junction compensation and PT100 current-drive bridges.
A thermocouple develops a Seebeck voltage from the temperature difference between its hot and cold junctions; type K gives about 41 microvolts per degree Celsius, but it measures hot relative to the cold junction at the terminal, so any cold-junction drift corrupts the reading. Place a temperature sensor such as an LM35 right at the terminal, add its voltage in the op amp or MCU as cold-junction compensation, and amplify the signal with an instrumentation amplifier by 100 to 200 times before filtering to the ADC. Use proper thermocouple extension wire and watch the polarity; a reversed connection flips the sign.
A PT100 reads 100 ohms at 0 degree Celsius and changes about 0.385 ohm per degree. Drive it with a 1mA op-amp current source so its drop starts at 100mV and rises with temperature. In three-wire hookup the two leads go to the differential inputs of the amplifier so lead resistance cancels symmetrically. Use an instrumentation amplifier such as the INA128 with gain G equals 1 plus 50k over Rg; choose Rg so a 200 degree span maps to 3.3V. Self-heating at 1mA is about 0.1mW and negligible, but at 5mA it raises the reading.
On site, keep the cold-junction sensor tight to the terminal or you measure the cabinet and lose 2 to 3 degrees in winter. Use three or four-wire for PT100 over distance; two-wire with 10 meters of lead adds about 2 ohms, some 5 degrees of error. Pick a low-bias instrumentation amplifier so input bias current does not drop voltage across the leads, and use a low-drift sense resistor in the current source. Calibrate at two points, ice water at 0 and boiling water at 100, not just one trim.
| Sensor | Signal | Op amp handling |
|---|---|---|
| Type K TC | 41uV/°C, needs CJC | in-amp ×100-200 + CJC |
| PT100 two-wire | 100Ω@0°C, 0.385Ω/°C | 1mA source + in-amp |
| PT100 three-wire | lead resistance cancels | diff in-amp symmetric |
| Range | K -200 to 1300°C | PT100 -200 to 600°C |
| Self-heating | ≤1mA negligible | 5mA clearly high |
| Bias effect | low-bias in-amp | large bias drops voltage |
| Calibration | 0/100°C two points | ice and boiling water |
Thermocouple and PT100 front ends serve furnace, pipe and bearing temperature monitoring. Do cold-junction compensation with the sensor at the terminal, use three-wire PT100 to kill lead error, pick a low-drift sense resistor, set the in-amp gain to the span, and calibrate at ice and boiling water.
The quantities and component characteristics discussed here follow current national standards and the corresponding IEC documents: Temperature front-end design per GB/T 4728; platinum RTD per IEC 60751, thermocouple per IEC 60584. Always check the device datasheet and the product design file for ratings and derating.
Locate the issue with the approach above, then confirm with measurements; safety-related circuits must follow the standards and manufacturer requirements.